beautypg.com

Relay board, Motherboard, A/d functions – Teledyne 6200E - Sulfides Analyzer User Manual

Page 255

background image

Model 6200E Instruction Manual

TROUBLESHOOTING & REPAIR

M6200E Rev: A1

255

You can verify this failure by logging on to the instrument using APICOM or a terminal program. If
the analyzer responds to remote commands and the display changes accordingly, the display
wiring or the I

2

C bus may be faulty.

11.5.7. Relay Board

The relay board circuit can most easily be checked by observing the condition of its status LEDs as
described in Section 11.1.4, and the associated output when toggled on and off through the

SIGNAL I/O function in the DIAG menu, see Section 6.9.2.

If the front panel display responds to key presses and D1 on the relay board is not flashing, then
either the I2c connection between the motherboard and the relay board is bad, or the relay board
itself is bad.

If D1 on the relay board is flashing, but toggling an output in the

Signal I/O function menu does

not toggle the output’s status LED, the there is a circuit problem, or possibly a blown driver chip,
on the relay board.

If D1 on the Relay board is flashing and the status indicator for the output in question (heater,
valve, etc.) toggles properly using the

Signal I/O function, but the output device does not turn

on/off, then the associated device (valve or heater) or its control device (valve driver, heater
relay) is malfunctioning.

Several of the control devices are in sockets and can easily be replaced. The table below lists the
control device associated with a particular function:

Table 11-6: Relay Board Control Devices

FUNCTION CONTROL

DEVICE

SOCKETED

Valve0 – Valve3

U5

Yes

Valve4 – Valve7

U6

Yes

All heaters

K1-K5

Yes

11.5.8. Motherboard

11.5.8.1. A/D functions

A basic check of the analog to digital (A/D) converter operation on the motherboard is to use the

Signal I/O function under the DIAG menu. Check the following two A/D reference voltages and
input signals that can be easily measured with a voltmeter.

• Using the Signal I/O function (Section 6.9.2 and Appendix D), view the value of

REF_4096_MV and REF_GND. If these signals are within 10 mV and 3 mV, respectively,

of their nominal values (4096 and 0) and are stable to within ±0.5 mV, the basic A/D
converter is functioning properly. If these values fluctuate largely or are off by more than

specified above, one or more of the analog circuits may be overloaded or the motherboard
may be faulty.

• Choose one parameter in the Signal I/O function such as SAMPLE_PRESSURE (see

previous section on how to measure it). Compare its actual voltage with the voltage
displayed through the

SIGNAL I/O function. If the wiring is intact but there is a difference

of more than ±10 mV between the measured and displayed voltage, the motherboard may
be faulty.