7.4. Manual Calibration with Zero/Span Valves

Zero and Span calibrations using the Zero/Span Valve option are similar to that described in
Section 7.2, except that:

Zero air and span gas is supplied to the analyzer through the zero gas and span gas inlets rather

than through the sample inlet.

The zero and cal operations are initiated directly and independently with dedicated keys (CALZ &
CALS
)

STEP ONE: Connect the sources of zero air and span gas to the respective ports on the rear panel
(Figure 3-2) as shown below.

External Zero

Air Scrubber

Filter

MODEL 701

Zero Air

Generator

Source of

SAMPLE Gas

VENT if input is pressurized

MODEL 700

Gas Dilution Calibrator

(with O

3

generator option)

VENT


MODEL

6200E

Sample

Zero Air

Exhaust

Span

Calibrated H

2

S gas

(At high concentration)

Figure 7-2: Setup for Manual Calibration with Z/S Valve Option Installed