beautypg.com

Manually configuring the network ip addresses – Teledyne 6200E - Sulfides Analyzer User Manual

Page 110

background image

Operating Instructions

Model 6200E Instruction Manual

110

M6200E Rev: A1

6.10.6.3. Manually Configuring the Network IP Addresses

There are several circumstances when you may need to manually configure the interface settings
of the analyzer’s Ethernet card. The INET sub-menu may also be used to edit the Ethernet card’s

configuration properties

• Your LAN is not running a DHCP software package,

• The DHCP software is unable to initialize the analyzer’s interface;

• You wish to program the interface with a specific set of IP addresses that may not be the

ones automatically chosen by DHCP.

Editing the Ethernet Interface properties is a two step process.

STEP 1: Turn DHCP OFF: While DHCP is turned ON, the ability to manually set INSTRUMENT IP,
GATEWAY IP
and SUBNET MASK is disabled


ENTR accept

new settings

EXIT ignores

new settings

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST TST > CAL

SETUP

SETUP X.X

COMMUNICATIONS MENU

ID

INET

COM1

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS DIAG

EXIT

SETUP X.X

DHCP: ON

ON

ENTR EXIT

SETUP X.X

DHCP: ON

EDIT

EXIT

SETUP X.X

DHCP: ON

OFF

ENTR

EXIT

Continue with editing of Ethernet interface

properties (see Step 2, below).