Teledyne 6200E - Sulfides Analyzer User Manual
Page 109
Model 6200E Instruction Manual
Operating Instructions
M6200E Rev: A1
109
NOTE
It is a good idea to check these settings the first time you power up your analyzer after it
has been physically connected to the LAN/Internet to make sure that the DHCP has
successfully downloaded the appropriate information from you network server(s).
If the gateway IP, instrument IP and the subnet mask are all zeroes (e.g. “0.0.0.0”), the
DCHP was not successful.
You may have to manually configure the analyzer’s Ethernet properties.
See your network administrator.
To view the above properties, press:
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG
EXIT
SETUP X.X
DHCP: ON
SET>
EDIT
EXIT
SETUP X.X
INST IP: 0.0.0.0
EXIT
SETUP X.X
GATEWAY IP: 0.0.0.0
EXIT
SETUP X.X
SUBNET MASK: 0.0.0.0
EXIT
SETUP X.X
TCP PORT: 3000
EDIT
EXIT
SETUP X.X
HOSTNAME: M101E
EDIT
EXIT
EDIT Key
Disabled
From this point on,
EXIT returns to
COMMUNICATIONS
MENU
Don not alter unless
directed to by Teledyne
Instruments Customer
Service personnel