Operating Instructions

Model 6200E Instruction Manual

72

M6200E Rev: A1

• Single range (SNGL) mode: This mode sets a single maximum range for the analog output.

If single range is selected (see Section 6.7.4) both outputs are slaved together and will

represent the same measurement span (e.g. 0-50 ppm), however their electronic signal
levels may be configured differently (e.g. 0-10 VDC vs. 0-.1 VDC – See Section 6.9.4.1). In

SO2/H2S multigas measurement mode, although the two inputs are measuring different
gases, the two measurements scales are identical

Independent range (IND) mode: This mode allows the A1 and A2 outputs to be configured
with different measurement spans (see Section 6.7.5) as well as separate electronic signal

levels (see Section 6.9.4.1) and, if the instrument is equipped with the SO

2

/H

2

S multigas

measurement option, different gas measurements.

• Auto range (AUTO) mode: As in single range mode, both outputs are slaved together and

will represent the same measurement span, however this mode gives the analyzer the
ability switch to automatically switch between the two user selected ranges (High and Low).

This switching occurs dynamically as the concentration value fluctuates.

High/low range status is output via the External Digital I/O Status Bits (see Section
6.12.1.1).

To select the Analog Output Range Type press:

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP X.X

RANGE MODE:

SNGL

SNGL

IND AUTO

ENTR EXIT

EXIT Returns

to the Main

SAMPLE Display

Go To

Section

6.7.4

Go To

Section

6.7.5

Go To

Section

6.7.6

Only one of the

range modes may

be active at any

time.