6.11.2.11. HOLDOFF Feature

The iDAS HOLDOFF feature allows to prevent data collection during calibrations and during the
DAS_HOLDOFF period enabled and specified in the VARS (Section 6.8). To enable or disable the
HOLDOFF, follow the instruction shown in section 6.11.2.2 then press

:


Edit Data Channel Menu

From the DATA ACQUISITION menu

SETUP X.X

0) CONC: ATIMER, 2, 4032, R


PREV NEXT INS DEL

EDIT

PRNT

EXIT

Exits to the main

Data Acquisition

menu

SETUP X.X

NAME:CONC


SET>

EDIT

PRINT

EXIT

SETUP X.X

CAL HOLD OFF:ON


SET>

EDIT

PRINT

EXIT

ENTR accepts the new

setting and returns to the

previous menu.

EXIT ignores the new setting

and returns to the previous

menu.

SETUP X.X

CAL HOLD OFF:ON

ON

ENTR

EXIT

Toggle key to turn

HOLDOFF ON or OFF

Press SET> key until…