6.11.2.10. Disabling/Enabling Data Channels

Data channels can be temporarily disabled, which can reduce the read/write wear on the disk-on-
chip. The ALL_01 channel of the M6200E, for example, is disabled by default.

To disable a data channel, follow the instruction shown in section 6.11.2.2 then press:


Edit Data Channel Menu

From the DATA ACQUISITION menu

SETUP X.X

0) CONC: ATIMER, 2, 4032, R


PREV NEXT INS DEL

EDIT

PRNT

EXIT

Exits to the main

Data Acquisition

menu

SETUP X.X

NAME:CONC


SET>

EDIT

PRINT

EXIT

SETUP X.X

CHANNEL ENABLE:ON


EDIT

PRINT

EXIT

ENTR accepts the new

setting and returns to the

previous menu.

EXIT ignores the new setting

and returns to the previous

menu.

SETUP X.X

CHANNEL ENABLE:ON

OFF

ENTR EXIT

Toggle key to turn

channel ON or OFF

Press SET> key until…