Table 3-7: Possible Warning Messages at Start-Up

MESSAGE MEANING

ANALOG CAL WARNING

The instruments A/D circuitry or one of its analog outputs is not

calibrated.

AZERO WRN XXX.X MV

The reading taken during the auto-zero cycle is outside of specified limits.
The value XXX.X indicates the auto-zero reading at the time of the

warning.

BOX TEMP WARNING

The temperature inside the M6200E chassis is outside the specified limits.

CANNOT DYN SPAN

Remote span calibration failed while the dynamic span feature was set to
active

CANNOT DYN ZERO

Remote zero calibration failed while the dynamic zero feature was set to
active

CONFIG INITIALIZED

Configuration was reset to factory defaults or was erased.

DARK CAL WARNING

Dark offset above limit specified indicating that too much stray light is
present in the sample chamber.

DATA INITIALIZED

iDAS data storage was erased.

FRONT PANEL WARN

Firmware is unable to communicate with the front panel.

HVPS WARNING

High voltage power supply for the PMT is outside of specified limits.

IZS TEMP WARNING

On units with IZS options installed: The permeation tube temperature is

outside of specified limits.

PMT DET WARNING

PMT detector output outside of operational limits.

PMT TEMP WARNING

PMT temperature is outside of specified limits.

RCELL PRESS WARN

Sample chamber pressure is outside of specified limits.

RCELL TEMP WARNING

Sample chamber temperature is outside of specified limits.

REAR BOARD NOT DET

The CPU is unable to communicate with the motherboard.

RELAY BOARD WARN

The firmware is unable to communicate with the relay board.

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

SAMPLE PRESS WARN

Sample pressure outside of operational parameters.

SYSTEM RESET

The computer was rebooted.

UV LAMP WARNING

The UV lamp intensity measured by the reference detector reading too
low or too high

To view and clear warning messages:

SAMPLE

HVPS WARNING SO2 = 0.00

TEST

CAL

MSG

CLR

SETUP

Make sure warning messages are

not due to real problems.

Press CLR to clear the current

message.

If more than one warning is active, the

next message will take its place

Once the last warning has been

cleared, the analyzer returns to

SAMPLE mode

SAMPLE

RANGE=500.000 PPB

SO2 = 0.00


< TST TST > CAL

MSG

CLR

SETUP

SAMPLE

SYSTEM RESET

SO2 = 0.00


TEST CAL

MSG

CLR

SETUP

TEST deactivates warning

messages

MSG activates warning

messages.

> keys replaced with

TEST key

NOTE:

If the warning message persists

after several attempts to clear it,

the message may indicate a

real problem and not an artifact

of the warm-up period