Figure 6-6: setup for calibrating current outputs – Teledyne 6200E - Sulfides Analyzer User Manual

Page 92

background image

Operating Instructions

Model 6200E Instruction Manual

92

M6200E Rev: A1

to output and from instrument to instrument, you will need to measure the change in the current
with a current meter placed in series with the output circuit (Figure 6-6).

mA

IN OUT

V OUT +

V OUT -

I IN +

I IN -

Recording

Device

Analyzer

See Table 3-2 for

pin assignments of

the Analog Out

connector on the

rear panel.

Figure 6-6: Setup for Calibrating Current Outputs

NOTE

Do not exceed 60 V between current loop outputs and instrument ground.

To adjust the zero and span values of the current outputs, activate the ANALOG I/O

CONFIGURATION MENU (see Section 6.9.1), then press

: