Operating Instructions

Model 6200E Instruction Manual

138

M6200E Rev: A1

Figure 6-16: iDAS Configuration Through a Terminal Emulation Program.

Both procedures are best started by downloading the default iDAS configuration, getting familiar
with its command structure and syntax conventions, and then altering a copy of the original file
offline before uploading the new configuration.

CAUTION

Whereas the editing, adding and deleting of iDAS channels and parameters of one

channel through the front-panel keyboard can be done without affecting the other

channels, uploading an iDAS configuration script to the analyzer through its

communication ports will erase all data, parameters and channels by replacing them

with the new iDAS configuration. Backup of data and the original iDAS configuration is

advised before attempting any iDAS changes.