Teledyne 6200E - Sulfides Analyzer User Manual
Page 79
Model 6200E Instruction Manual
Operating Instructions
M6200E Rev: A1
79
To access and navigate the VARS menu, use the following key sequence.
See Section 6.8.1
for instructions
regarding this setting
SETUP X.X
0 ) MEASURE_MODE=H2S
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
1 ) DAS_HOLD_OFF=
15.0 Minutes
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
1 ) TPC_ENABLE=ON
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
DAS_HOLD_OFF=15.0 Minutes
1 5 .0
ENTR EXIT
SETUP X.X
3)RCELL_SET=
50.0 DegC
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
3) IZS_SET=
50.0 DegC
PREV
NEXT
JUMP
EDIT
PRNT EXIT
DO NOT change
theses set-points
unless
specifically
instructed to by
T-API Customer
Service.
SETUP X.X
CLOCK_ADJ=0 Sec/Day
+ 0 0
ENTR EXIT
SETUP X.X
8) CLOCK_ADJ=0 Sec/Day
PREV
NEXT
JUMP
EDIT
PRNT EXIT
EXIT ignores the new setting.
ENTR accepts the new setting.
SETUP X.X
TPC_ENABLE=ON
ON
ENTR EXIT
SETUP X.X
5 ) DYN_ZERO=ON
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
DYN_ZERO=ON
ON
ENTR EXIT
Toggle this keys to change setting
Toggle this keys to change setting
SETUP X.X
6) DYN_SPAN=ON
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
DYN_SPAN=ON
ON
ENTR EXIT
Toggle this keys to change setting
Toggle this keys to change setting
Toggle tese keys to change setting
SETUP X.X
CONC_PRECUISION : 3
AUTO 0 1 2 3 4 ENTR EXIT
SETUP X.X
7) CONC_PRECUISION : 1
PREV
NEXT
JUMP
EDIT
PRNT EXIT
Toggle these keys to change setting
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP