APPENDIX A-4: M6200E Signal I/O Definitions, Revision A.1
Model 6200E Instruction Manual
292
M6200E Rev A.1
APPENDIX A-4: M6200E Signal I/O Definitions, Revision A.1
Table A-4: M6200E Signal I/O Definitions, Revision A.1
SIGNAL NAME
BIT OR CHANNEL
NUMBER
DESCRIPTION
Internal outputs, U8, J108, pins 1–8 = bits 0–7, default I/O address 322 hex
ELEC_TEST
0
1 = electrical test on
0 = off
OPTIC_TEST
1
1 = optic test on
0 = off
PREAMP_RANGE_HI
2
1 = select high preamp range
0 = select low range
3–5 Spare
I2C_RESET
6
1 = reset I
2
C peripherals
0 = normal
I2C_DRV_RST
7
0 = hardware reset 8584 chip
1 = normal
Control inputs, U11, J1004, pins 1–6 = bits 0–5, default I/O address 321 hex
EXT_ZERO_CAL
0
0 = go into zero calibration
1 = exit zero calibration
EXT_SPAN_CAL
1
0 = go into span calibration
1 = exit span calibration
EXT_LOW_SPAN
2
2
0 = go into low span calibration
1 = exit low span calibration
3–5
Spare
6–7
Always
1
Control inputs, U14, J1006, pins 1–6 = bits 0–5, default I/O address 325 hex
0–5
Spare
6–7
Always
1
Control outputs, U17, J1008, pins 1–8 = bits 0–7, default I/O address 321 hex
0–7
Spare
Control outputs, U21, J1008, pins 9–12 = bits 0–3, default I/O address 325 hex
0–3
Spare
Alarm outputs, U21, J1009, pins 1–12 = bits 4–7, default I/O address 325 hex
ST_SYSTEM_OK2
1
4
1 = system OK
0 = any alarm condition
5–7
Spare
A status outputs, U24, J1017, pins 1–8 = bits 0–7, default I/O address 323 hex
ST_SYSTEM_OK
0
0 = system OK
1 = any alarm condition
ST_CONC_VALID
1
0 = conc. valid
1 = hold off or other conditions
ST_HIGH_RANGE
2
0 = high auto-range in use
1 = low auto-range