Model 6200E Instruction Manual

Theory Of Operation

M6200E Rev: A1

225

adjustments should only be performed when encountering problems with the software calibration
that cannot be rectified otherwise. See Section 11.6.3 for this hardware calibration.

To

Motherboard

PMT Preamp PCA

Low Pass

Noise

Filter

E Test Control

From CPU

MUX

Amp to

Voltage

Converter/

Amplifier

D-A

Converter

PMT Coarse

Gain Set

(Rotary

Switch)

PMT Fine

Gain Set

(Rotary

Switch)

PMT

Signal

Offset

E-Test

Generator

O-Test

Generator

O Test Control

From CPU

PMT Temp

Sensor

PMT

Temperature

Feedback

Circuit

TEC Control

PCA

PMT Output

PMT HVPS

Drive Voltage

PMT Temp Analog Signal

to Motherboard

PMT Output Signal

(PMT) to Motherboard

O Test

LED

Figure 10-16: PMT Preamp Block Diagram

The PMT temperature control loop maintains the PMT temperature around 7° C and can be viewed
as test function PMT TEMP on the front panel.

The electrical test (ETEST) circuit generates a constant, electronic signal intended to simulate the
output of the PMT (after conversion from current to voltage). By bypassing the detector’s actual
signal, it is possible to test most of the signal handling and conditioning circuitry on the PMT
preamplifier board. See section 6.9.6 for instructions on performing this test.

The optical test (OTEST) feature causes an LED inside the PMT cold block to create a light signal
that can be measured with the PMT. If zero air is supplied to the analyzer, the entire
measurement capability of the sensor module can be tested including the PMT and the current to
voltage conversion circuit on the PMT preamplifier board. See section 6.9.5 for instructions on
performing this test.