Model 6200E Instruction Manual

Getting Started

M6200E Rev: A1

27

3.1.1.4. Connecting the Serial Ports

If you wish to utilize either of the analyzer’s two serial interfaces, refer to Section 6.10 and 6.12
of this manual for instructions on configuration and usage.

3.1.1.5. Connecting to a LAN or the Internet

If your unit has a Teledyne Analytical Instruments Ethernet card (Option 63), plug one end of the
7’ CAT5 cable supplied with the option into the appropriate place on the back of the analyzer (see

Figure 5-6 in Section 5.6.3) and the other end into any nearby Ethernet access port.

3.1.1.6. Connecting to a LAN or the Internet

If your unit has a Teledyne Analytical Instruments RS-232 multidrop card (Option 62), see
section 10.6.7 fo instructions on setting it up.

3.1.2. Pneumatic Connections:

CAUTION

To prevent dust from getting into the analyzer, it was shipped with small plugs inserted

into each of the pneumatic fittings on the rear panel. Make sure that all dust plugs are

removed before attaching exhaust and supply gas lines.

Sample and calibration gases should only come into contact with PTFE (Teflon) or glass materials.
They should not come in contact with FEP or stainless steel materials.

Figures 3-6 and 3-7 show the most common configurations for gas supply and exhaust lines to the
Model 6200E Analyzer. Figure 3-8 shows the connections for units with valve options installed.

Please refer to Figure 3-2 for pneumatic connections at the rear panel and Table 3-4 for
nomenclature.

Table 3-4: Inlet / Outlet Connector Nomenclature

REAR PANEL LABEL

FUNCTION

SAMPLE

Connects the sample gas to the analyzer. When operating the analyzer without
zero span option, this is also the inlet for any calibration gases.

EXHAUST

Connects the exhaust of the analyzer with the external vacuum pump.

SPAN

On units with zero/span/shutoff valve options installed, connect a gas line to
the source of calibrated span gas here.

ZERO AIR

On Units with zero/span valve or IZS option installed, this port connects the

zero air gas or the zero air cartridge to the analyzer.