Model 6200E Instruction Manual

Operating Instructions

M6200E Rev: A1

123

PNUMTC: Collects sample flow and sample pressure data at five minute intervals and stores an
average once a day with a time and date stamp. This data is useful for monitoring the condition of

the pump and critical flow orifice (sample flow) and the sample filter (clogging indicated by a drop
in sample pressure) over time to predict when maintenance will be required. The last 360 daily

averages (about 1 year) are stored.

CALDAT: Logs new slope and offset every time a zero or span calibration is performed. This Data

Channel also records the instrument reading just prior to performing a calibration. Note: this Data
Channel collects data based on an event (a calibration) rather than a timer. This Data Channel will

store data from the last 200 calibrations. This does not represent any specific length of time since
it is dependent on how often calibrations are performed. As with all Data Channels, a time and date
stamp is recorded for every data point logged.

The Channel Properties. Triggering Events and Data Parameters/Functions for these default
channels are: