6.11.2.3. Trigger Events

To edit the list of data parameters associated with a specific data channel, press:


SETUP X.X

EVENT:ATIMER

ENTR

EXIT

Edit Data Channel Menu

SETUP X.X

0) CONC: ATIMER, 2, 4032,R


PREV NEXT INS DEL

EDIT

PRNT

EXIT

Exits to the Main

Data Acquisition

menu

Press each key repeatedly to cycle through the

list of available trigger events.

SETUP X.X

NAME:CONC


SET>

EDIT

PRINT

EXIT

SETUP X.X

EVENT:ATIMER


EDIT

PRINT

EXIT

ENTR accepts the new string

and returns to the previous

menu.

EXIT ignores the new string

and returns to the previous

menu.

From the DATA ACQUISITION menu

(see Section 6.11.2.2)

6.11.2.4. Editing iDAS Parameters

Data channels can be edited individually from the front panel without affecting other data channels.
However, when editing a data channel, such as during adding, deleting or editing parameters, all

data for that particular channel will be lost, because the iDAS can store only data of one format
(number of parameter columns etc.) for any given channel. In addition, an iDAS configuration can

only be uploaded remotely as an entire set of channels. Hence, remote update of the iDAS will
always delete all current channels and stored data.

To modify, add or delete a parameter, follow the instruction shown in section 6.11.2.2 then press: