Lamp calibration – Teledyne 6200E - Sulfides Analyzer User Manual
Page 97
Model 6200E Instruction Manual
Operating Instructions
M6200E Rev: A1
97
6.9.7. Lamp Calibration
An important factor in accurately determining H
2
S concentration, once the H
2
S is converter to SO
2,
is the amount of UV light available to transform the SO
2
into SO
2
* (see Section 10.1.1 and
10.1.2). The model 6200E compensates for variations in the intensity of the available UV light by
adjusting the H
2
S concentration calculation using a ratio (LAMP RATIO)that results from dividing
the current UV lamp (UV LAMP) intensity by a value stored in the CPU’s memory (LAMP_CAL).
Both LAMP Ration and UV Lamp are test functions viewable from the instruments front panel.
To cause the analyzer to measure and record a value for LAMP_CAL, press.
Exit at
any time
to return
to main
the
SETUP
menu
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Repeat Pressing NEXT until . .
.
DIAG
LAMP CALIBRATION
PREV NEXT
ENTR
EXIT
DIAG FCAL
LAMP CAL VALUE:4262.4 mV
4 2 6 2 .4 ENTR EXIT
ENTR accepts the
new value
EXIT ignores the new
value
The value displayed is the
current output of the UV
source reference detector
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT