Model 6200E Instruction Manual

Operating Instructions

M6200E Rev: A1

97

6.9.7. Lamp Calibration

An important factor in accurately determining H

2

S concentration, once the H

2

S is converter to SO

2,

is the amount of UV light available to transform the SO

2

into SO

2

* (see Section 10.1.1 and

10.1.2). The model 6200E compensates for variations in the intensity of the available UV light by

adjusting the H

2

S concentration calculation using a ratio (LAMP RATIO)that results from dividing

the current UV lamp (UV LAMP) intensity by a value stored in the CPU’s memory (LAMP_CAL).

Both LAMP Ration and UV Lamp are test functions viewable from the instruments front panel.

To cause the analyzer to measure and record a value for LAMP_CAL, press.

Exit at

any time
to return

to main

the

SETUP

menu

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

Repeat Pressing NEXT until . .

.

DIAG

LAMP CALIBRATION

PREV NEXT

ENTR

EXIT

DIAG FCAL

LAMP CAL VALUE:4262.4 mV


4 2 6 2 .4 ENTR EXIT

ENTR accepts the

new value

EXIT ignores the new

value

The value displayed is the

current output of the UV

source reference detector

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT