beautypg.com

Frequently asked questions & glossary, Faq’s – Teledyne 6200E - Sulfides Analyzer User Manual

Page 41

background image

Model 6200E Instruction Manual

Frequently Asked Questions & Glossary

M6200E Rev: A1

41

4. FREQUENTLY ASKED QUESTIONS & GLOSSARY

4.1. FAQ’s

The following list contains some of the most commonly asked questions relating to the Model

6200E H

2

S Analyzer.

Q: Why is the ZERO or SPAN key not displayed during calibration?
A: The M6200E disables certain keys whenever the chosen value is out of range for that particular

parameter. In this case, the expected span or zero value is too different from the actually

measured value for the calibration process to be valid.

Q: Why does the ENTR key sometimes disappear on the front panel display?
A: Sometimes the ENTR key will disappear if you select a setting that is invalid or out of the

allowable range for that parameter, such as trying to set the 24-hour clock to 25:00:00 or a
range to more than 20 000 ppb. Once you adjust the setting to an allowable value, the ENTR

key will re-appear.

Q: Can I automate the calibration of my analyzer?
A: Any analyzer with zero/span valve or IZS option can be automatically calibrated using the

instrument’s AutoCal feature.
However, the accuracy of the IZS option’s permeation tube is ±5%. While this may be

acceptable for basic calibration checks, the IZS option is not permitted as a calibration source

in applications following US EPA protocols.
To achieve highest accuracy, it is recommended to use cylinders of calibrated span gases in
combination with a zero air source. Teledyne Analytical Instruments offers a zero air generator

Model 701 and a gas dilution calibrator Model 700 for this purpose.

Q: What do I do if the concentration on the instrument's front panel display does not match the

value recorded or displayed on my data logger even if both instrument's are properly
calibrated?

A: This most commonly occurs for one or both of the following reasons: (1) a difference in circuit

ground between the analyzer the data logger, or (2) a scale problem with the input to the data

logger. The analog outputs of the analyzer can be manually calibrated to compensate for

either or both of these effects, see Section 6.9.4.2 & 6.9.4.3.

Q: How do I measure the sample flow?
A: Sample flow is measured by attaching a calibrated flow meter to the sample inlet port when

the instrument is operating. The sample flow should be 650 cm³/min ±10%. Section 11.5.2

includes detailed instructions on performing a check of the sample gas flow.

Q: How often do I need to change the particulate filter?
A: Once per week. Table 9-1 contains a maintenance schedule listing the most important, regular

maintenance tasks.