beautypg.com

Available analog output signals, Figure 6-4: analog output connector key – Teledyne 6200E - Sulfides Analyzer User Manual

Page 70

background image

Operating Instructions

Model 6200E Instruction Manual

70

M6200E Rev: A1

6.7. SETUP – RNGE: Analog Output Reporting Range

Configuration

6.7.1. Available Analog Output Signals

The analyzer has three active analog output signals, accessible through a connector on the rear

panel.

ANALOG OUT

A1 A2 A3 A4
+ - + - + - + -

SO

2

concentration

outputs

HIGH range when

DUAL mode is selected

Not Used

Test Channel

LOW range when

DUAL mode is selected

Figure 6-4: Analog Output Connector Key

All three outputs can be configured either at the factory or by the user for full scale outputs of 0.1
VDC, 1VDC, 5VDC or 10VDC. Additionally A1 and A2 may be equipped with optional 0-20 mADC
current loop drivers and configured for any current output within that range (e.g. 0-20, 2-20, 4-20,

etc.). The user may also adjust the signal level and scaling of the actual output voltage or current
to match the input requirements of the recorder or datalogger (See Section 6.9.4.3 & 6.9.4.5).

In its basic configuration, the A1 and A2 channels of the M6200E output a signal that is
proportional to the H

2

S concentration of the sample gas. Several operating modes are available

which allow them to be slaved together (SNGL Mode, see Section 6.7.4 or AUTO mode, se section
6.7.6) or operate independently (IND mode, see Section 6.7.5) The user may also select between
a variety of reporting range spans as well:

EXAMPLE:

A1 OUTPUT: Output Signal = 0-5 VDC representing 0-1000 ppm concentration values

A2 OUTPUT: Output Signal = 0 – 10 VDC representing 0-500 ppm concentration values.