beautypg.com

Dilution ratio – Teledyne 6200E - Sulfides Analyzer User Manual

Page 77

background image

Model 6200E Instruction Manual

Operating Instructions

M6200E Rev: A1

77

6.7.8. Dilution Ratio

The dilution ratio is a software option that allows the user to compensate for any dilution of the

sample gas before it enters the sample inlet. Using the dilution ratio option is a 4-step process:

1. Select reporting range units: Follow the procedure in Section 6.7.7.

2. Select the range: Use the procedures in Section 6.7.3 – 6.7.6. Make sure that the SPAN value

entered is the maximum expected concentration of the undiluted calibration gas and that the
span gas is either supplied through the same dilution inlet system as the sample gas or has an

appropriately lower actual concentration. For example, with a dilution set to 100, a 1 ppm gas
can be used to calibrate a 100 ppm sample gas if the span gas is not routed through the
dilution system. On the other hand, if a 100 ppm span gas is used, it needs to pass through the
same dilution steps as the sample gas.

3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts dilutent and 1 part of

sample gas):

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST TST > CAL

SETUP

SETUP C.3

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP C.3

RANGE CONTROL MENU

MODE

SET UNIT

DIL

EXIT

DIL only appears

if the dilution ratio

option has been

installed

SETUP C.3

DIL FACTOR: 1.0 GAIN

0 0 0 1 .0

ENTR

EXIT

Toggle these keys to set the dilution

factor.

This is the number by which the

analyzer will multiply the H

2

S

concentrations of the gas passing

through the reaction cell.

SETUP C.3

DIL FACTOR: 20.0 GAIN

0 0 2 0 .0

ENTR

EXIT

EXIT ignores the

new setting.

ENTR accepts the

new setting.

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

The analyzer multiplies the measured gas concentrations with this dilution factor and displays the
result.

NOTE

Once the above settings have been entered, the instrument needs to be recalibrated

using one of the methods discussed in Chapter 7.