Instrument Maintenance

Model 6200E Instruction Manual

202

M6200E Rev: A1

5. If there is a PMT response to the external light, symmetrically tighten the sample chamber

mounting screws or replace the 1/4” vacuum tubing with new, black PTFE tubing (this tubing

will fade with time and become transparent). Often, light leaks are also caused by O-rings
being left out of the assembly.

6. Carefully replace the analyzer cover.

7. If tubing or O-rings were changed, carry out a leak check (Section 11.5.1).

User Notes: