beautypg.com

Teledyne 6200E - Sulfides Analyzer User Manual

Page 297

background image

Model 6200E Instruction Manual

APPENDIX A-5: M6200E iDAS Functions, Revision A.1

M6200E Rev A.1

297

Table A-6: M6200E iDAS Functions, Revision A.1

NAME DESCRIPTION

UNITS

PMTDET

PMT detector reading

mV

UVDET

UV lamp intensity reading

mV

LAMPR

UV lamp ratio of calibrated intensity

%

DRKPMT

PMT electrical offset

mV

DARKUV

UV lamp electrical offset

mV

S2SLP1

SO

2

slope for range #1

PPB/mV

S2SLP2

SO

2

slope for range #2

PPB/mV

H2SLP1 or TRSLP1 H

2

S/TRS slope for range #1

PPB/mV

H2SLP2 or TRSLP2 H

2

S/TRS slope for range #2

PPB/mV

S2OFS1

SO

2

offset for range #1

mV

S2OFS2

SO

2

offset for range #2

mV

H2OFS1 or TROFS1 H

2

S/TRS offset for range #1

mV

H2OFS2 or TROFS2 H

2

S/TRS offset for range #2

mV

S2ZSC1

SO

2

concentration for range #1 during zero/span calibration, just

before computing new slope and offset

PPB

S2ZSC2

SO

2

concentration for range #2 during zero/span calibration, just

before computing new slope and offset

PPB

H2ZSC1 or TRZSC1 H

2

S/TRS concentration for range #1 during zero/span calibration,

just before computing new slope and offset

PPB

H2ZSC2 or TRZSC2 H

2

S/TRS concentration for range #2 during zero/span calibration,

just before computing new slope and offset

PPB

S2CNC1

SO

2

concentration for range #1

PPB

S2CNC2

SO

2

concentration for range #2

PPB

H2CNC1 or TRCNC1 H

2

S/TRS concentration for range #1

PPB

H2CNC2 or TRCNC2 H

2

S/TRS concentration for range #2

PPB

STABIL

SO

2

or H

2

S/TRS concentration stability (select with STABIL_GAS

variable)

PPB

STRLGT

Stray light reading

PPB

RCTEMP

Reaction cell temperature

°C

IZSTMP

IZS temperature

°C

PMTTMP

PMT temperature

°C

CNVTMP

1

Converter temperature

°C

SMPFLW

Sample flow

cc/m

SMPPRS

Sample pressure

“Hg

BOXTMP

Internal box temperature

°C

HVPS

High voltage power supply output

Volts

TEST8

Diagnostic test input (TEST_INPUT_8)

mV

TEST11

Diagnostic test input (TEST_INPUT_11)

mV

TEMP4

Diagnostic temperature input (TEMP_INPUT_4)

°C

TEMP5

Diagnostic temperature input (TEMP_INPUT_5)

°C

TEMP6

Diagnostic temperature input (TEMP_INPUT_6)

°C

REFGND

Ground reference (REF_GND)

mV

RF4096

4096 mV reference (REF_4096_MV)

mV