Model 6200E Instruction Manual
Optional Hardware and Software
M6200E Rev: A1
51
CAUTION
Gas flow must be maintained at all time for units with a permeation tube installed.
Insufficient gas flow allows gas to build up to levels that will contaminate the
instrument.
5.4.3. IZS Permeation Tubes (Options 53, 55 & 57)
Several replacement permeation tubes are available for the IZS option. They are identical in size
and shape but are designed to have different effusion rates. Since the M6200E can be configured
as either a H
2
S or SO
2
analyzer, both types of permeation tubes are included here.
OPTION GAS
TYPE
EFFUSION RATE
(± 25%)
APPROXIMATE
CONCENTRATION
SPECIFIED FLOW
RATE
OPT 52
H
2
S
106 ng/min
800 – 1200 ppb
0.76 lpm
OPT 53
SO
2
421 ng/min
300 - 500 ppb
0.76 lpm
Opt 54
H
2
S
53 ng/min
400 – 600 ppb
0.76 lpm
OPT 55
SO
2
842
ng/min
600 - 1000 ppb
0.76 lpm
OPT 57
SO
2
222 ng/min
800 – 1200 ppb
0.56 lpm
Each tube comes with a calibration certificate, traceable to a NIST standard, specifying its actual
effusion rate of that tube to within ±5% when immersed in a gas stream moving at the specified
flow rate. This calibration is performed at a tube temperature of 50° C.
NOTE
The instrument can only be fitted with one type of permeation tube at a time. Therefore
the IZS option can only be used to calibrate or check the instrument for one gas, H
2
S or
SO
2
, but not both.
5.4.4. Zero Air Scrubber Maintenance Kit (Option 43)
This kit includes the following items needed to refurbish the zero air scrubber, included with the
IZS valve options.
TELEDYNE ANALYTICAL
INSTRUMENTS PART NO.
DESCRIPTION
005960000 Activated
charcoal
FL0000001
Sintered filter for span gas inlet fitting
1
FL0000003
Replacement particulate filter
OR0000001 O-Ring
(qty:2)
1
Formatted: Bullets and Numbering