Model 6200E Instruction Manual

Operating Instructions

M6200E Rev: A1

85

The following DC current output limits apply to the current loop modules:

Table 6-10: Analog Output Current Loop Range

RANGE MINIMUM

OUTPUT

MAXIMUM

OUTPUT

0-20 mA

0 mA

20 mA

These are the physical limits of the current loop modules, typical applications use 2-20 or 4-20
mA for the lower and upper limits. Please specify desired range when ordering this option.
The default offset for all ranges is 0 mA.

Pin assignments for the output connector at the rear panel of the instrument are shown in Table 6-

11.

ANALOG OUT

A1 A2 A3 A4
+ - + - + - + -

Table 6-11: Analog Output Pin Assignments

PIN ANALOG

OUTPUT

VOLTAGE

SIGNAL

CURRENT

SIGNAL

1

V Out

I Out +

2

A1

Ground I

Out

-

3

V Out

I Out +

4

A2

Ground I

Out

-

5

A3

Not Used

Not Used

7

V Out

not available

8

A4

Ground not

available

See Figure 3-2 for a the location of the analog output connector on the instruments rear panel.