Dte and dce communication – Teledyne 6200E - Sulfides Analyzer User Manual

Page 106

background image

Operating Instructions

Model 6200E Instruction Manual

106

M6200E Rev: A1

When COM2 is configured for RS-485 operation the port uses the same female DB-9 connector on

the back of the instrument as when Com2 is configured for RS-232 operation, however, the pin
assignments are different.

Female DB-9 (COM2)

(As seen from outside analyzer)









(RS-485)

1

2 3

4

5

6

7

8

9

GND

RX/TX

+

RX/TX

-

Figure 6-10: Back Panel connector Pin-Outs for COM2 in RS-485 mode.

The signal from this connector is routed from the motherboard via a wiring harness to a 6-pin

connectors on the CPU card, CN5.

CN5

(Located on CPU card)










(As seen from inside analyzer)

2

4

6

1

3

5

RX/TX

+

RX/TX

-

GND

Figure 6-11: CPU connector Pin-Outs for COM2 in RS-485 mode.

6.10.5. DTE and DCE Communication

RS-232 was developed for allowing communications between data terminal equipment (DTE) and
data communication equipment (DCE). Basic terminals always fall into the DTE category whereas
modems are always considered DCE devices. The difference between the two is the pin assignment

of the Data Receive and Data Transmit functions.

• DTE devices receive data on pin 2 and transmit data on pin 3.

• DCE devices receive data on pin 3 and transmit data on pin 2.