beautypg.com

Teledyne 6200E - Sulfides Analyzer User Manual

Page 180

background image

EPA Protocol Calibration

Model 6200E Instruction Manual

180

M6200E Rev: A1

sampling and through as much of the ambient air inlet system as is practicable. If the Instrument
will be used on more than one range (i.e. IND or AUTO ranges), it should be calibrated

separately on each applicable range. Calibration documentation should be maintained with each
analyzer and also in a central backup file.

Table 8-1: Activity Matrix for Calibration Equipment & Supplies

EQUIPMENT/SUPPLIES

ACCEPTANCE LIMITS

FREQUENCY AND

METHOD OF

MEASUREMENT

ACTION IF

REQUIREMENTS ARE

NOT MET

Recorder

Compatible with output
signal of analyzer; min.
chart width of 150 mm (6

in) is recommended

Check upon receipt Return equipment to

supplier

Sample Line And

Manifold

Constructed of PTFE or
glass

Check upon receipt Return equipment to

supplier

Calibration Equipment Meets guidelines of

reference 1 and Section
2.3.2 (Q.A. Handbook)

See Section 2.3.9
(Q.A. Handbook)

Return equipment/
supplies to supplier or
take corrective action

Working Standard SO

2

Cylinder Gas Or SO

2

Permeation Tube

Traceable to NIST-SRM
meets limits in traceability
protocol for accuracy and
stability (Section 2.0.7,

Q.A. Handbook)

Analyzed against
NIST-SRM; see
protocol in Section
2.0.7, Q.A.

Handbook

Obtain new working
standard and check for
traceability

Zero Air

Clean dry ambient air, free
of contaminants that cause
detectable response with
the SO

2

analyzer.

See Section 2.9.2
(Q.A. Handbook)

Obtain air from another
source or regenerate.

Record Form

Develop standard forms

N/A

Revise forms as
appropriate

Audit Equipment

Must not be the same as
used for calibration

System must be
checked out against
known standards

Locate problem and
correct or return to
supplier

Table 8-2: Activity Matrix for Calibration Procedure

EQUIPMENT/

SUPPLIES

ACCEPTANCE LIMITS

FREQUENCY AND METHOD OF

MEASUREMENT

ACTION IF REQUIREMENTS ARE

NOT MET

Calibration

Gases

NIST traceable

Assayed against an NIST-SRM
semi-annually, Sec. 2.0.7,
(Q.A. Handbook)

Working gas standard is
unstable, and/or measurement
method is out of control; take

corrective action such as
obtaining new calibration gas.

Dilution Gas Zero air, free of

contaminants

See Section 2.9.2 (Q.A.
Manual)

Return to supplier or take
appropriate action with
generation system

Multi-Point
Calibration

Use calibration
procedure in Subsec.
2.2 (Q.A. Handbook);
also Federal Register

Perform at least once every
quarter or anytime a level span
check indicates a discrepancy,
or after maintenance which
may affect the calibration;
Subsec 2.5 (Q.A. Manual)

Repeat the calibration