Teledyne 6200E - Sulfides Analyzer User Manual
Page 180

EPA Protocol Calibration
Model 6200E Instruction Manual
180
M6200E Rev: A1
sampling and through as much of the ambient air inlet system as is practicable. If the Instrument
will be used on more than one range (i.e. IND or AUTO ranges), it should be calibrated
separately on each applicable range. Calibration documentation should be maintained with each
analyzer and also in a central backup file.
Table 8-1: Activity Matrix for Calibration Equipment & Supplies
EQUIPMENT/SUPPLIES
ACCEPTANCE LIMITS
FREQUENCY AND
METHOD OF
MEASUREMENT
ACTION IF
REQUIREMENTS ARE
NOT MET
Recorder
Compatible with output
signal of analyzer; min.
chart width of 150 mm (6
in) is recommended
Check upon receipt Return equipment to
supplier
Sample Line And
Manifold
Constructed of PTFE or
glass
Check upon receipt Return equipment to
supplier
Calibration Equipment Meets guidelines of
reference 1 and Section
2.3.2 (Q.A. Handbook)
See Section 2.3.9
(Q.A. Handbook)
Return equipment/
supplies to supplier or
take corrective action
Working Standard SO
2
Cylinder Gas Or SO
2
Permeation Tube
Traceable to NIST-SRM
meets limits in traceability
protocol for accuracy and
stability (Section 2.0.7,
Q.A. Handbook)
Analyzed against
NIST-SRM; see
protocol in Section
2.0.7, Q.A.
Handbook
Obtain new working
standard and check for
traceability
Zero Air
Clean dry ambient air, free
of contaminants that cause
detectable response with
the SO
2
analyzer.
See Section 2.9.2
(Q.A. Handbook)
Obtain air from another
source or regenerate.
Record Form
Develop standard forms
N/A
Revise forms as
appropriate
Audit Equipment
Must not be the same as
used for calibration
System must be
checked out against
known standards
Locate problem and
correct or return to
supplier
Table 8-2: Activity Matrix for Calibration Procedure
EQUIPMENT/
SUPPLIES
ACCEPTANCE LIMITS
FREQUENCY AND METHOD OF
MEASUREMENT
ACTION IF REQUIREMENTS ARE
NOT MET
Calibration
Gases
NIST traceable
Assayed against an NIST-SRM
semi-annually, Sec. 2.0.7,
(Q.A. Handbook)
Working gas standard is
unstable, and/or measurement
method is out of control; take
corrective action such as
obtaining new calibration gas.
Dilution Gas Zero air, free of
contaminants
See Section 2.9.2 (Q.A.
Manual)
Return to supplier or take
appropriate action with
generation system
Multi-Point
Calibration
Use calibration
procedure in Subsec.
2.2 (Q.A. Handbook);
also Federal Register
Perform at least once every
quarter or anytime a level span
check indicates a discrepancy,
or after maintenance which
may affect the calibration;
Subsec 2.5 (Q.A. Manual)
Repeat the calibration