Functional check – Teledyne 6200E - Sulfides Analyzer User Manual

Page 36

background image

Getting Started

Model 6200E Instruction Manual

36

M6200E Rev: A1

3.2.4. Functional Check

1. After the analyzer’s components have warmed up for at least 30 minutes, verify that the

software properly supports any hardware options that were installed.

2. Check to make sure that the analyzer is functioning within allowable operating parameters.

Appendix C includes a list of test functions viewable from the analyzer’s front panel as well as
their expected values. These functions are also useful tools for diagnosing performance

problems with your analyzer (Section 11.1.2). The enclosed Final Test and Validation Data

sheet (part number 04551) lists these values before the instrument left the factory.

To view the current values of these parameters press the following key sequence on the

analyzer’s front panel. Remember until the unit has completed its warm up these parameters
may not have stabilized.

1

Only appears if IZS option is

installed.

2

Only appears if analog output A4

is actively reporting a test function.

3

Shown as they appear when analyzer

is in H

2

S mode. In SO

2

mode appear as SO2 STB, SO2 OFFS &

SO2 SLOPE. In multigas mode, both versions appear.

RANGE

H2S STB

3

PRES
SAMP FL
PMT

NORM PMT
UV LAMP
LAMP RATIO
STR. LGT

DARK PMT
DARK LAMP
H2S SLOPE

3

H2S OFFS

3

HVPS
RCELL TEMP
BOX TEMP

PMT TEMP
IZS TEMP

1

TEST

2

TIME

SAMPLE

RANGE = 500.0 PPB

NOX = X.X


< TST

TST >

CAL

SETUP

Refer to

Section

6.2.1 for

definitions

of these

test

functions.

Toggle keys to

scroll through list of functions

3. If your analyzer has an Ethernet card (Option 63) installed and your network is running a

dynamic host configuration protocol (DHCP) software package, the Ethernet option will
automatically configure its interface with your LAN. However, it is a good idea to check these

settings to make sure that the DHCP has successfully downloaded the appropriate network
settings from your network server (See Section 6.10.6.2).


If your network is not running DHCP, you will have to configure the analyzer’s interface

manually (See Section 6.10.6.3).

NOTE

Once you have completed the above set-up procedures, please fill out the Quality

Questionnaire that was shipped with your unit and return it to Teledyne Analytical

Instruments.

This information is vital to our efforts in continuously improving our service and our

products.

THANK YOU.