Functional check – Teledyne 6200E - Sulfides Analyzer User Manual
Page 36

Getting Started
Model 6200E Instruction Manual
36
M6200E Rev: A1
3.2.4. Functional Check
1. After the analyzer’s components have warmed up for at least 30 minutes, verify that the
software properly supports any hardware options that were installed.
2. Check to make sure that the analyzer is functioning within allowable operating parameters.
Appendix C includes a list of test functions viewable from the analyzer’s front panel as well as
their expected values. These functions are also useful tools for diagnosing performance
problems with your analyzer (Section 11.1.2). The enclosed Final Test and Validation Data
sheet (part number 04551) lists these values before the instrument left the factory.
To view the current values of these parameters press the following key sequence on the
analyzer’s front panel. Remember until the unit has completed its warm up these parameters
may not have stabilized.
1
Only appears if IZS option is
installed.
2
Only appears if analog output A4
is actively reporting a test function.
3
Shown as they appear when analyzer
is in H
2
S mode. In SO
2
mode appear as SO2 STB, SO2 OFFS &
SO2 SLOPE. In multigas mode, both versions appear.
RANGE
H2S STB
3
PRES
SAMP FL
PMT
NORM PMT
UV LAMP
LAMP RATIO
STR. LGT
DARK PMT
DARK LAMP
H2S SLOPE
3
H2S OFFS
3
HVPS
RCELL TEMP
BOX TEMP
PMT TEMP
IZS TEMP
1
TEST
2
TIME
SAMPLE
RANGE = 500.0 PPB
NOX = X.X
< TST
TST >
CAL
SETUP
Refer to
Section
6.2.1 for
definitions
of these
test
functions.
Toggle
scroll through list of functions
3. If your analyzer has an Ethernet card (Option 63) installed and your network is running a
dynamic host configuration protocol (DHCP) software package, the Ethernet option will
automatically configure its interface with your LAN. However, it is a good idea to check these
settings to make sure that the DHCP has successfully downloaded the appropriate network
settings from your network server (See Section 6.10.6.2).
If your network is not running DHCP, you will have to configure the analyzer’s interface
manually (See Section 6.10.6.3).
NOTE
Once you have completed the above set-up procedures, please fill out the Quality
Questionnaire that was shipped with your unit and return it to Teledyne Analytical
Instruments.
This information is vital to our efforts in continuously improving our service and our
products.
THANK YOU.