Calibration Procedures

Model 6200E Instruction Manual

160

M6200E Rev: A1

STEP ONE: Connect the sources of zero air and span gas as shown below.

No Valve Options Installed

MODEL

6200E

Sample

Zero Air

Exhaust

Span

MODEL 701

Zero Air

Generator

VENT

Calibrated H

2

S GAS

(At high concentration)

Source of

SAMPLE Gas

Removed

during

Calibration

Valve

OR

Needle valve to control flow

MODEL

6200E

Sample

Zero Air

Exhaust

Span

MODEL 701

Zero Air

Generator

VENT

Calibrated H

2

S GAS

(At span gas concentration)

Source of

SAMPLE Gas

Removed

during

calibration

MODEL 700 Gas

Dilution

Calibrator

(with Ozone Bench

Option)

Figure 7-1: Setup for Manual Calibration without Z/S valve or IZS Option