Model 6200E Instruction Manual

Theory Of Operation

M6200E Rev: A1

237

Instrument slope and offset values recorded during the last calibration can be viewed by pressing
the following keystroke sequence

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X


< TST TST > CAL

SETUP

SAMPLE TIME = HH:MM:SS

H2S

=XXX.X


< TST TST > CAL

SETUP

SAMPLE PMT TEMP=0.0C

H2S

=XXX.X


< TST TST > CAL

SETUP

SAMPLE BOX TEMP=0.0C

H2S

=XXX.X


< TST TST > CAL

SETUP

SAMPLE RCELL TEMP=0.0C

H2S

=XXX.X


< TST TST > CAL

SETUP

SAMPLE

HVPS 553 VOLTS

H2S

=XXX.X


< TST TST > CAL

SETUP

SAMPLE H2S OFFS=XX.X MV

H2S

=XXX.X


< TST TST > CAL

SETUP

SAMPLE

H2S SLOPE=XXX

H2S

=XXX.X


< TST TST > CAL

SETUP

NOTE

Separate slope and offset values are calculated and recorded for H2S and SO2 gas

measurements. Here they are shown as they appear when analyzer is in H2S mode.

In SO2 mode appear as SO2 OFFS & SO2 SLOPE. In multigas mode, both versions

appear.

10.7.3. Temperature and Pressure Compensation (TPC)

Feature

As explained previously, changes in temperature can significantly effect the amount of fluoresced
UV light generated in the instruments sample chamber. To negate this effect the Model 6200E

maintains the sample gas at a stable, raised temperature.

Pressure changes can also have a noticeable, if more subtle, effect on the H

2

S concentration

calculation. To account for this, the Model 6200E software includes a feature which allows the
instrument to include a compensation factor in the H

2

S calculations that is based on changes in

ambient pressure.

When the TPC feature is enabled, the analyzer’s H

2

S concentration is divided by a factor called

PRESSCO which is based on the ratio between the ambient pressure of the sample gas and

standard atmospheric pressure (Equation 10-5). As ambient pressure increases, the compensated
H

2

S concentration is decreased.

A)

-

HG

("

29.92

_SLOPE

SAMP_PRESS

A)

-

HG

("

SSURE

SAMPLE_PRE

PRESSCO

×

=

Equation 10-5

SAMPLE-PRESSURE: The ambient pressure of the sample gas as measured by the instrument’s

sample pressure sensor (see Figure 10-7) in “Hg-A.

SAMP_PRESS_SLOPE: Sample pressure slope correction factor. The default setting for

Section 6.8 describes the method for enabling/disabling the TPC feature.