beautypg.com

Extended warranty (options 92 & 93), Special software features, Maintenance mode switch – Teledyne 6200E - Sulfides Analyzer User Manual

Page 56: Second language switch, Dilution ratio option

background image

Optional Hardware and Software

Model 6200E Instruction Manual

56

M6200E Rev: A1

5.8. Extended Warranty (Options 92 & 93)

Two options are available for extending Teledyne Analytical Instruments’ standard warranty
(Section 2.3). Both options have to be specified upon ordering the analyzer.

Option Number

Description

OPT 92

Extends warranty to cover a two (2) year period from the date of
purchase.

OPT 93

Extends warranty to cover a five (5) year period from the date of
purchase.

5.9. Special Software Features

5.9.1. Maintenance Mode Switch

Teledyne Analytical Instruments analyzers are equipped with a switch that places the instrument in

maintenance mode. When present, the switch is accessed by opening the hinged front panel and is
located on the rearward facing side of the display/keyboard driver PCA; on the left side; near the
particulate filter.

When in maintenance mode the instrument ignores all commands received via the COMM ports that

alter the operation state of the instrument This includes all calibration commands, diagnostic menu
commands and the reset instrument command. The instrument continues to measure

concentration and send data when requested.

This feature is of particular use for instruments connected to multidrop or Hessen protocol
networks.

5.9.2. Second Language Switch

Teledyne Analytical Instruments analyzers are equipped with a switch that activates an alternate

set of display messages in a language other than the instrument’s default language. This switch is
accessed by opening the hinged front panel and is located on the rearward facing side of the

display/keyboard driver PCA; on the right side.

To activate this feature, the instrument must also have a specially programmed Disk on Chip
containing the second language. Contact Teledyne Analytical Instruments Customer Service

personnel for more information.

5.9.3. Dilution Ratio Option

The dilution ration feature is a software option that is designed for applications where the sample

gas is diluted before being analyzed by the Model 6200E. Typically this occurs in continuous
emission monitoring (CEM) applications where the quality of gas in a smoke stack is being tested
and the sampling method used to remove the gas from the stack dilutes the gas.