beautypg.com

Calibration quality – Teledyne 6200E - Sulfides Analyzer User Manual

Page 177

background image

Model 6200E Instruction Manual

Calibration Procedures

M6200E Rev: A1

177

If the instrument is in multigas measurement mode and the SNGL reporting range mode is
selected, the AutoCal sequence is programmed as described above. If the reporting range of the

M6200E must is set for either IND or AUTO. the gas type and range to be measured must be
specified. This parameter appears at the end of the programming sequences after the

CALIBRATE: ON/OFF parameter is set. For example:

SETUP X.X

GAS-RANGE: H2S-HIGH


EDIT

EXIT

SETUP X.X

CALIBRATE: ON


SET>

EDIT

EXIT

Follow standard AutoCal programming process to this

point, then …

SETUP X.X

GAS TO CAL: H2S

SO2

H2S

ENT

R EXIT

SETUP X.X

Gas TO CAL: SO2

LOW HIGH

ENTR

EXIT

EXIT returns

to the SETUP

Menu

SETUP X.X

SEQ 2) ZERO–SPAN, 2:00:30


PREV NEXT

MODE

SET

EXIT

Sequence

Delta Time

MODE Delta Days

SETUP X.X

RANGE TO CAL: LOW

LOW

HIGH

ENT

R EXIT

SETUP X.X

RANGE TO CAL: HIGH

LOW HIGH

ENTR

EXIT

SETUP X.X

GAS-RANGE: SO2-LOW


EDIT

EXIT

7.10. Calibration Quality

After completing one of the calibration procedures described above, it is important to evaluate the
analyzer’s calibration slope and offset parameters. These values describe the linear response curve
of the analyzer. The values for these terms, both individually and relative to each other, indicate

the quality of the calibration. To perform this quality evaluation, you will need to record the values
of both test functions (Section 6.2.1 or Appendix A-3), all of which are automatically stored in the
iDAS channel CALDAT for data analysis, documentation and archival.

Make sure that these parameters are within the limits listed in Table 7-5.

Table 7-5: Calibration Data Quality Evaluation

FUNCTION

MINIMUM VALUE

OPTIMUM VALUE

MAXIMUM VALUE

H2S SLOPE

1

-0.700 1.000 1.300

H2S OFFS

1

50.0 mV

n/a

250.0 mV

These values should not be significantly different from the values recorded on the Teledyne
Analytical Instruments Final Test and Validation Data sheet that was shipped with your instrument.

If they are, refer to the troubleshooting Chapter 11.

1

Shown as they appear when analyzer is in H2S mode. In SO2 mode they appear as

SO2 OFFS

&

SO2 SLOPE. In multigas mode, both version are listed and should be checked,