beautypg.com

Cleaning or changing critical flow orifices – Teledyne 6200E - Sulfides Analyzer User Manual

Page 200

background image

Instrument Maintenance

Model 6200E Instruction Manual

200

M6200E Rev: A1

4. The sample chamber will separate into two halves, the stainless steel manifold assembly and

the black plastic sample chamber with window, stainless steel cylinder and O-rings.

5. The sample chamber (both plastic part and stainless steel cylinder) and glass window should

be cleaned with de-ionized water and a clean lint free cloth and dried thereafter.

6. Usually it is not necessary to clean the sample and ozone flow orifices since they are protected

by sintered filters. If tests show that cleaning is necessary, refer to section 9.3.7. below on
how to clean/replace the critical flow orifice.

7. Do not remove the sample and ozone nozzles. They are Teflon threaded and require a special

tool for reassembly. If necessary, the manifold with nozzles attached can be cleaned in an

ultrasonic bath.

8. Reassemble in proper order and re-attach onto sensor housing. Reconnect pneumatics and

heater connections, then re-attach the pneumatic sensor assembly and the cleaning procedure
is complete.

9.3.7. Cleaning or Changing Critical Flow Orifices

A critical flow orifice, located on the exhaust manifold maintains the proper flow rate of gas
through the M6200E analyzer. Refer to section 10.3.3 for a detailed description of its functionality

and location. Despite the fact this these device is protected by sintered stainless steel filters, it
can, on occasion, clog, particularly if the instrument is operated without a sample filter or in an

environment with very fine, sub-micron particle-size dust.

The M6200E introduces an orifice holder that makes changing the orifice very easy. In fact, it is
recommended to keep spare orifice holder assemblies at hand to minimize downtime and swap
orifices in a matter of a few minutes. Appendix B lists several complete spare part kits for this
purpose. To clean or replace a critical flow orifice, do the following:

• Turn off power to the instrument and vacuum pump. Remove the analyzer cover and

locate the reaction cell (Section 3.1 and Figure 3-9).

• Unscrew the 1/8” sample and ozone air tubes from the reaction cell

• For orifices on the reaction cell: Unscrew the orifice holder with a 9/16” wrench. This part

holds all components of the critical flow assembly as shown in Figure 9-4. Appendix B
contains a list of spare part numbers.

• For orifices in the vacuum manifold: the assembly is similar to the one shown in Figure 9-

4, but without the orifice holder, part number 04090, and bottom O-ring OR34 and with an
NPT fitting in place of the FT 10 fitting. After taking off the connecting tube, unscrew the
NPT fitting.

• Take out the components of the assembly: a spring, a sintered filter, two O-rings and the

orifice. For the vacuum manifold only, you may need to use a scribe or pressure from the
vacuum port to get the parts out of the manifold.

• Discard the two O-rings and the sintered filter and clean the critical flow orifice in a

ultrasonic cleaner for about 30 minutes using methanol. Let the part dry.

• Re-assemble the parts as shown in Figure 9-4 and re-connect them to the reaction cell

manifold or the vacuum manifold.