10 Sensors

916 Ti-Touch

■■■■■■■■

83

10.5

Limit values for the calibration data

Sensor: Edit

Limit values

You can define the following limit values in the dialog Edit sensor /
Limit values
:

Slope (pH electrodes)

Electrode zero point (pH electrodes)

These values are monitored during the calibration. If these limits are
infringed, then a message will be displayed and you can decide whether
to accept the calibration data or not.

Monitoring slope

on | off (Default value: off)

If this parameter is activated, then the slope will be monitored.

Lower limit

pH electrodes:
Input range

–999.9 - 999.9 %

Default value

96.0 %

Upper limit

pH electrodes:
Input range

–999.9 - 999.9 %

Default value

101.0 %