beautypg.com

2 titration parameters, Titration parameters – Metrohm 916 Ti-Touch User Manual

Page 295

background image

■■■■■■■■■■■■■■■■■■■■■■

28 Parameters

916 Ti-Touch

■■■■■■■■

279

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.

Input range

0 - 999999 s

Default value

1 s

28.1.2

Titration parameters

Under [Titration parameters], the parameters influencing the sequence
of the entire titration are defined.

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

slow | optimal | fast | user

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted equivalence points.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.
fast
For fast and less critical titrations.
user
The individual titration parameters can be modified.

NOTE

Select optimal as titration rate when you are developing a new titra-
tion method. This parameter is suitable for virtually all titrations and
only needs adaptation in special cases.

The settings of the individual titration rates are listed in table 6, page
282
.

Temperature

Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement
= automatic or continuous is defined (see sensor dialog