Metrohm 916 Ti-Touch User Manual
Page 298
28.1 Dynamic equivalence point titrations (DET)
■■■■■■■■■■■■■■■■■■■■■■
282
■■■■■■■■
916 Ti-Touch
Measuring mode Upol:
Input range
0.01 - 99.90 µA/min
Default value
50.00 µA/min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed. This can be useful when the titration reaction pro-
ceeds slowly or the electrode is slow to respond.
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.
Input range
0 - 999999 s
Default value
0 s
Max. waiting time
If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.
Input range
0 - 999999 s
Default value
26 s
Table 6
Default values of the predefined titration rates for DET
Titration rate
slow
optimal
fast
Meas. point density
2
4
6
Min. increment
10.00 µL
10.00 µL
30.00 µL
Max. increment
off
off
off
Dosing rate
maximum
maximum
maximum
Signal drift
– pH, U and Ipol
– Upol
20.0 mV/min
20.0 µA/min
50.0 mV/min
50.0 µA/min
80.0 mV/min
80.0 µA/min
Min. waiting time
0 s
0 s
0 s
Max. waiting time
38 s
26 s
21 s