28.1 Dynamic equivalence point titrations (DET)

■■■■■■■■■■■■■■■■■■■■■■

282

■■■■■■■■

916 Ti-Touch

Measuring mode Upol:
Input range

0.01 - 99.90 µA/min

Default value

50.00 µA/min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed. This can be useful when the titration reaction pro-
ceeds slowly or the electrode is slow to respond.

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.

Input range

0 - 999999 s

Default value

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.

Input range

0 - 999999 s

Default value

26 s

Table 6

Default values of the predefined titration rates for DET

Titration rate

slow

optimal

fast

Meas. point density

2

4

6

Min. increment

10.00 µL

10.00 µL

30.00 µL

Max. increment

off

off

off

Dosing rate

maximum

maximum

maximum

Signal drift

– pH, U and Ipol

– Upol

20.0 mV/min

20.0 µA/min

50.0 mV/min

50.0 µA/min

80.0 mV/min

80.0 µA/min

Min. waiting time

0 s

0 s

0 s

Max. waiting time

38 s

26 s

21 s