beautypg.com

12 communication, 1 scanning remote lines (scan) – Metrohm 916 Ti-Touch User Manual

Page 399

background image

■■■■■■■■■■■■■■■■■■■■■■

28 Parameters

916 Ti-Touch

■■■■■■■■

383

the selected titrant, the validity of the titer, the working life of the titrant
and the time interval for the GLP test of the buret unit are checked at the
start of the determination.

Selection

Selection of configured titrants | not defined

Default value

not defined

not defined
No check takes place.

28.12 Communication

The following communications commands are available:

Scanning remote
lines (SCAN)

Defining remote signals which are awaited
before the next method command is started.

Setting remote
lines (CTRL)

Defining remote signals.

Receiving RS-232
commands (SCAN
RS)

Defining RS-232 commands which are awaited
before the next method command is started.

Sending RS-232
commands (CTRL
RS)

Defining the RS-232 commands which are sent.

28.12.1 Scanning remote lines (SCAN)

The command SCAN can be used to define input signals at the remote
interface which are awaited before the next method command is started.

Control device

This parameter will only be shown when, in addition to the Ti-Touch, a
Sample Processor has been configured as control device. Control devices
are defined under System

Device manager.

Selection of the control device from the list of devices.

Selection

Selection of configured control devices

Remote Box

Selection of the MSB connector the Remote Box is connected to. The
selection does not depend on the number of MSB connectors of the con-
trol device. The remote signal defined is requested on this Remote Box.

Selection

1 | 2 | 3 | 4

Default value

1