2 setting remote lines (ctrl) – Metrohm 916 Ti-Touch User Manual
Page 401
■■■■■■■■■■■■■■■■■■■■■■
28 Parameters
916 Ti-Touch
■■■■■■■■
385
Document message
In the determination data it will be documented, that the time interval
has been expired.
Cancel determination
The determination is stopped.
28.12.2 Setting remote lines (CTRL)
The command CTRL can be used to define output signals at the remote
interface.
Control device
This parameter will only be shown when, in addition to the Ti-Touch, a
Sample Processor has been configured as control device. Control devices
are defined under System
▶ Device manager.
Selection of the control device from the list of devices.
Selection
Selection of configured control devices
Remote Box
Selection of the MSB connector the Remote Box is connected to. The
selection does not depend on the number of MSB connectors of the con-
trol device. The remote signal defined is sent via this Remote Box.
Selection
1 | 2 | 3 | 4
Default value
1
Output signal
Selection of the signal out of the templates or entering the required bit
pattern. Templates are defined under System
▶ Templates ▶ Output
lines.
NOTE
A line set active is not being reset automatically, not even at the end of
the determination.
Entering a bit pattern:
■
0 = line inactive
■
1 = line active
■
* = retain line status
■
p = set pulse
The output lines are always numbered from right to left, i.e. with the sig-
nal *************1 line 0 is set. With a pulse, the length is set to 200 ms.