28 Parameters

916 Ti-Touch

■■■■■■■■

299

Measuring mode pH:
Input range

–20.000 - 20.000

Selection

off

Default value

off

Measuring mode U, Ipol:
Input range

–2,000.0 - 2,000.0 mV

Selection

off

Default value

off

Measuring mode Upol:
Input range

–200.0 - 200.0 µA

Selection

off

Default value

off

Stop EP

The titration is stopped when the specified number of equivalence points
has been found.

Input range

1 - 9

Default value

9

Selection

off

Volume after EP

This volume will be added when the number of equivalence points
defined under Stop EP has been found. The curve shape after the equiva-
lence point can also be seen this way.

Input range

0.01000 - 9999.99 mL

Selection

off

Default value

off

Stop time

The titration is stopped when the specified time has elapsed since the start
of the titration.

Input range

0 - 999999 s

Selection

off

Default value

off

Filling rate

Rate at which the dosing cylinder is filled.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum