28 Parameters
916 Ti-Touch
■■■■■■■■
311
fast
For flat titration curves for which the endpoint is reached only slowly.
user
The individual titration parameters can be modified.
The settings of the individual titration rates are listed in table 8, page
312.
User-defined parameters
These parameters are only accessible when Titration rate = user.
Dynamics
This parameter defines the control range before the specified endpoint.
Individual volume steps are dosed in the control range, the dosing is finely
controlled. The closer the endpoint, the slower the dosing until the dosing
rate defined under Min. rate has been reached. The larger the control
range, the slower the titration. Outside the control range, dosing is carried
out continuously, and the dosing rate is defined under Max. rate.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 429.
Measuring mode pH:
Input range
0.001 - 20.000
Default value
2.000
Selection
off
Measuring mode U and Ipol:
Input range
0.1 - 2,000.0 mV
Default value
100.0 mV
Selection
off
Measuring mode Upol:
Input range
0.01 - 200.00 µA
Default value
10.00 µA
Selection
off
Max. rate
Rate at which dosing is carried out outside of the control range.
Input range
0.01 - 166.00 mL/min
Default value
10.00 mL/min
Selection
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 406).