28.2 Monotonic equivalence point titrations (MET)

■■■■■■■■■■■■■■■■■■■■■■

300

■■■■■■■■

916 Ti-Touch

maximum
The maximum filling rate depends on the cylinder volume (see Chapter
31, page 406)
.

28.2.4

Potentiometric evaluation

The parameters for the evaluation of the titration curve are defined under
[Potentiometr. evaluation].

EP criterion

The equivalence point criterion found (ERC = Equivalence point Recogni-
tion Criterion) is compared with this value. Equivalence points whose ERC
is less than the value defined here will be ignored.

Measuring mode pH:
Input range

0.10 - 9.99

Default value

0.50

Measuring mode U, Ipol:
Input range

1 - 999 mV

Default value

30 mV

Measuring mode Upol:
Input range

0.1 - 99.9 µA

Default value

2.0 µA

EP recognition

This parameter allows you to filter out only the equivalence points that are
being sought.

Selection

all | greatest | last | ascending | descending |
off

Default value

all

all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.