28.2 Monotonic equivalence point titrations (MET)
■■■■■■■■■■■■■■■■■■■■■■
294
■■■■■■■■
916 Ti-Touch
Selection
off
Default value
off
Measuring mode U, Ipol:
Input range
0 - 999 mV/mL
Selection
off
Default value
off
Measuring mode Upol:
Input range
0 - 99 µA/mL
Selection
off
Default value
off
Dosing rate
Rate to be dosed with until the start measured value or the start slope is
reached.
Input range
0.01 - 166.00 mL/min
Default value
5.00 mL/min
Selection
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 406).
Initial measured value
The conditions for the entry of the first measured value in the measuring
point list are defined under [Initial meas. value].
Signal drift
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. For this value, a suitable
maximum waiting time is automatically calculated except you already have
modified this waiting time.
Measuring mode pH, U and Ipol:
Input range
0.1 - 999.0 mV/min
Selection
off
Default value
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Measuring mode Upol:
Input range
0.01 - 99.90 µA/min
Selection
off
Default value
off