28.2 Monotonic equivalence point titrations (MET)

■■■■■■■■■■■■■■■■■■■■■■

294

■■■■■■■■

916 Ti-Touch

Selection

off

Default value

off

Measuring mode U, Ipol:
Input range

0 - 999 mV/mL

Selection

off

Default value

off

Measuring mode Upol:
Input range

0 - 99 µA/mL

Selection

off

Default value

off

Dosing rate

Rate to be dosed with until the start measured value or the start slope is
reached.

Input range

0.01 - 166.00 mL/min

Default value

5.00 mL/min

Selection

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 406)
.

Initial measured value
The conditions for the entry of the first measured value in the measuring
point list are defined under [Initial meas. value].

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. For this value, a suitable
maximum waiting time is automatically calculated except you already have
modified this waiting time.

Measuring mode pH, U and Ipol:
Input range

0.1 - 999.0 mV/min

Selection

off

Default value

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Measuring mode Upol:
Input range

0.01 - 99.90 µA/min

Selection

off

Default value

off