2 exchange unit, 1 maximum dosing and filling rate, 2 default parameters for preparing (prep) – Metrohm 916 Ti-Touch User Manual
Page 423: 3 stirring rate, 2 exchange unit 31.2.1, Maximum dosing and filling rate, Default parameters for preparing (prep)
■■■■■■■■■■■■■■■■■■■■■■
31 Appendix
916 Ti-Touch
■■■■■■■■
407
31.2
Exchange unit
31.2.1
Maximum dosing and filling rate
The maximum dosing rate and maximum filling rate for the exchange unit
depend on the cylinder volume:
Cylinder volume
Maximum rate
1 mL
3.00 mL/min
5 mL
15.00 mL/min
10 mL
30.00 mL/min
20 mL
60.00 mL/min
50 mL
150.00 mL/min
Independent of the cylinder volume, values ranging from 0.01 to 166.00
mL/min can always be entered. When the function is carried out the rate
will be, if necessary, decreased automatically to the highest possible value.
31.2.2
Default parameters for preparing (PREP)
The PREP command is used to rinse and fill the cylinder and tubings of
the exchange unit air bubble-free. You should carry out this function
before the first determination or once per day.
If no titrant is selected in the command, preparing will be carried out with
the following default parameters:
■
The entire cylinder volume is dosed twice at the maximum dosing rate.
■
The cylinder is filled at the maximum filling rate.
31.3
Stirring rate
The stirring rate can be adjusted in steps of –15 to +15.
The approximate rotational speed can be calculated with the following
formula:
Rotational speed/min (r/min) = 125 · stirring rate
Example:
Stirring rate set: 8
Rotational speed in rpm = 125 · 8 = 1000