8 Titrants

916 Ti-Touch

■■■■■■■■

51

8 Titrants

Main dialog: System

Titrants

This chapter describes how you can create a list of titrants used in the sys-
tem. Titrants can be used in intelligent buret units or in non-intelligent
buret units. Intelligent buret units have a built-in data chip on which the
data for the titrant is stored. This data is automatically read out during
attachment and entered in the titrant list.

The titrant list can contain a maximum of 30 titrants. The following data is
specified for each titrant:

Designation

Cylinder volume

Type

IDU: dosing unit with integrated data chip
IEU: exchange unit with integrated data chip
EU: exchange unit without data chip

MSB connector of the dosing device/control device (only when
exchange unit/dosing unit is attached)

Titrants in exchange units/dosing units with integrated data chips are
depicted in green lettering.

The following titrant data is stored in the titrant list:

Name
Each titrant in the system is identified by its unambiguous name.

Concentration

Current titer

Working life