28 Parameters
916 Ti-Touch
■■■■■■■■
331
Electrode check
on | off (Default value: off)
For the following electrodes, an electrode check can be carried out:
■
pH electrodes
■
Metal electrodes
■
Ion-selective electrodes
A check is made that the electrode is properly connected and that no
short-circuit is present. The electrode check is carried out when this com-
mand is started. Note that this electrode check has nothing to do with the
electrode test (command ELT).
Temp. measurement
Type of temperature measurement.
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature is mea-
sured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
will be used (see dialog of the titration and measuring parameters).
off
The temperature will not be measured. The temperature entered man-
ually is used (see dialog of the titration and measuring parameters).
28.5.4
Stirrer
The parameters for the stirrer are edited under [Stirrer].
Stirrer
Selection of the MSB connector the stirrer is connected to. The selection
does not depend on the number of MSB connectors of the control device.
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 31.3, page 407. The optimum stirring
rate can be tested in the manual control.