Metrohm 916 Ti-Touch User Manual
Page 124
![background image](/manuals/586928/124/background.png)
11.6 Sample racks
■■■■■■■■■■■■■■■■■■■■■■
108
■■■■■■■■
916 Ti-Touch
You can edit the data of the selected rack in the dialog Sample rack /
Edit rack data.
Beaker radius samples
Actual radius of the sample vessels at the general sample positions of the
rack.
This beaker radius may not be less than the minimum beaker radius
defined in the tower properties (see "Min. beaker radius," page 100). If
the lift is to be moved to the work position, then these two values will be
compared with one another.
Input range
1.0 - 100.0 mm
Selection
off
Default value
off
off
No check takes place.
Beaker sensor
Each time a sample position is to be moved to with the MOVE command,
the beaker sensor checks whether a vessel is present. In the MOVE com-
mand, you define the action that takes place if the beaker sensor does not
detect a vessel at the position being moved to.
Selection
Tower | Robotic arm | off
Default value
off
Robotic arm
A Swing Head with beaker sensor must be mounted. In addition, a
suitable work position must be defined for the lift, so that the robotic
arm touches the sample vessel. The work position is moved to for the
purpose of beaker detection.
off
No check takes place.
1. Calibration pos.
Position of the first calibration solution for automatic calibration with a
USB Sample Processor (see "Definition of the calibration positions", page
414).
Input range
1 - "highest rack position"
It is imperative that the remaining buffers/standards
be placed on the rack positions directly following.
Selection
Special beaker 1…n | off
Default value
off