beautypg.com

2 control device, 3 sensor, Control device – Metrohm 916 Ti-Touch User Manual

Page 356: Sensor

background image

28.7 Electrode test for pH electrodes (ELT)

■■■■■■■■■■■■■■■■■■■■■■

340

■■■■■■■■

916 Ti-Touch

Selection

Standard | Gel | Non-aqueous | Custom

Default value

Standard

Standard
Electrodes which contain e.g. KCl solution as reference electrolyte.
Gel
Electrodes which contain Idrolyte as reference electrolyte.
Non-aqueous
Electrodes, which contain a non-aqueous reference electrolyte, e.g.
TEABr in ethylene glycol or LiCl in ethanol.
Custom
Under System

Templates a custom electrode type can be defined.

Temperature

Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement
= automatic or continuous is defined (see sensor dialog
of the command).

Input range

–20.0 - 150.0 °C

Default value

25.0 °C

28.7.2

Control device

The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System

Device

manager.

NOTE

This button will only be shown when, in addition to the Ti-Touch, a
Sample Processor has been configured as control device.

Control device

Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command.

Selection

Selection of configured control devices

28.7.3

Sensor

The parameters for the sensor are edited under [Sensor].

Measuring input

Selection of the measuring input the sensor is connected to. The selection
is not dependent on whether the control device has one or two measur-
ing interfaces.