21.3 Canceling determinations manually
■■■■■■■■■■■■■■■■■■■■■■
208
■■■■■■■■
916 Ti-Touch
Sequence of the sample series
■
After a determination is started, a check is made as to
– whether all of the devices required are connected and ready for
operation.
– whether all titrants and sensors are present.
– whether the monitoring intervals for titrants, sensors and com-
mon variables are still valid.
– if a sample assignment table is activated (see Chapter 15.1.2,
page 153):
whether the entered sample identification appears in the sample
assignment table. If this is the case, then the assigned method
will be loaded automatically and the actual determination will be
started. If a different method has been entered in the sample
table, then it will be ignored.
■
At the end of the determination,
– the line in the sample table is deleted.
– the next determination is started (if you are working with the
autostart function enabled (see Chapter 17, page 187)).
21.3
Canceling determinations manually
You can cancel a determination at any time with the fixed key [
]. The
command being carried out at that moment is canceled and no further
commands are executed.
If you are working with the autostart function activated (see Chapter 17,
page 187), the entire series will be canceled.
NOTE
If you do not wish to cancel the entire determination, but only a single
command, use the function [Cancel command] in the live display. This
is particularly useful if you want to cancel a titration because the
sought-after equivalence point has already been determined. All subse-
quent commands, e.g. calculations and print report, will nevertheless
still be carried out.