27 Manual control

916 Ti-Touch

■■■■■■■■

265

27.5

Remote

Manual control

Remote

With the function [Remote], you can define manual output signals to the
remote interface of a connected Remote Box. It is not possible to define
input signals; the current status of the input lines is, however, displayed.

Inputs

Current status of the input lines.

Outputs

Selection of the signal out of the templates or entering the required bit
pattern. Templates are defined under System

Templates Output

lines.

Entering the bit pattern:

0 = line inactive

1 = line active

* = retain line status

p = set pulse

The output lines are always numbered from right to left, i.e. with the sig-
nal 00000000000001 line 0 is set. With a pulse, the length is set to 200
ms. If you wish to set pulses with other lengths, you have to define a cor-
responding template.

Entry

Bit pattern containing exactly 14 characters
or a max. of 24 characters for the name of
the template

Default value

00000000000001

Selection

Selection of the templates defined