28 Parameters

916 Ti-Touch

■■■■■■■■

343

Table 11

Possible errors during the electrode test

Test criterion

Message

Remedy

Buffer pH 9:

The drift in stirred solution is > 1 mV.

General problem

Connect an electrode.

Replace the defective electrode
cable.

Clean the diaphragm (see leaflet
for the electrode).

Replace the electrode.

Buffer pH 9:

–10 mV < U(not stirred) < 10 mV

and

The sum of the drift values after 1, 2,
3 and 4 min is < 12 mV/min.

Short circuit

Replace the electrode.

A response time does not fulfill the
limit value for the rating Usable
electrode
.

Glass membrane / Diaphragm

Clean the diaphragm (see leaflet for
the electrode).

The pH values are not defined for all
the buffers at the temperature mea-
sured.

No buffer data

Repeat the electrode test at a tem-
perature at which the pH values of
all the buffers are defined.

All the slopes fulfill the limit values
for the rating Usable electrode.

and

Uoff is outside the preset limits.

Unsuitable reference electrode

Repeat the electrode test with a
suitable reference system.

Replace the contaminated refer-
ence electrolyte.

Select Electrode type = Cus-
tom
and adjust the limits for
Uoff.

The streaming potential is too high.

Contaminated diaphragm

Clean the diaphragm (see leaflet for
the electrode).

Two slopes do not fulfill the limit val-
ues for the rating Usable electrode.

Wrong buffer

Repeat the electrode test with buf-
fers pH 4, 7 and 9.

All the slopes do not fulfill the limit
values for the rating Usable elec-
trode
.

Partial short circuit

Check the temperature sensor.

Enter the correct calibration tem-
perature.