28.7 Electrode test for pH electrodes (ELT)

■■■■■■■■■■■■■■■■■■■■■■

342

■■■■■■■■

916 Ti-Touch

"+": counterclockwise rotation

"–": clockwise rotation

Input range

–15 - 15

Default value

8

28.7.5

Further information

Table 10

Limit values of the three electrode types

Rating

Excellent electrode

Good electrode

Usable electrode

Electrode type "Standard"

Streaming potential

2.5 mV

3.0 mV

4.0 mV

Drift

2.0 mV/s

2.5 mV/s

3.0 mV/s

Min. slope

96.5 %

96.0 %

95.0 %

Max. slope

101.0 %

102.0 %

103.0 %

Response time

45 s

50 s

60 s

Lower limit Uoff

–15 mV

Upper limit Uoff

15 mV

Electrode type "Gel"

Streaming potential

3.0 mV

3.5 mV

4.5 mV

Drift

2.5 mV/s

3.0 mV/s

4.0 mV/s

Min. slope

96.5 %

96.0 %

95.0 %

Max. slope

101.0 %

102.0 %

103.0 %

Response time

60 s

75 s

90 s

Lower limit Uoff

–20 mV

Upper limit Uoff

20 mV

Electrode type "Non-aqueous"

Streaming potential

3.0 mV

4.5 mV

6.0 mV

Drift

5.0 mV/s

7.0 mV/s

9.0 mV/s

Min. slope

88.0 %

80.0 %

70.0 %

Max. slope

120.0 %

130.0 %

140.0 %

Response time

60 s

75 s

90 s

Lower limit Uoff

–50 mV

Upper limit Uoff

100 mV